Controller | Cell Growth |
Operating mode | timer, continuous and program operation |
Display | Tablet PC |
Operation | touch screen |
Data memory size | 32 GB |
Auto restart after power failure | yes |
Ethernet interface | yes |
Single use reactor interface | yes |
Thermostat interface | yes |
Ext. pump interface | yes |
Filter heater | yes |
LED light panel | yes |
Stirring motor | yes |
Calibration guide tepmerature | yes |
Calibration guide pH | yes |
Calibration guide pO2 | yes |
pH - temperature compensation | yes |
Mode hose filling | yes |
Stirrer control mode | off, manual, profile, Cascade-pO2 control, chaotic |
Speed range | 1 - 2200 rpm |
Speed deviation >= 100 rpm | +/-2 % |
Speed deviation <= 100 rpm | +/-5 rpm |
Speed display | TFT |
Reversible direction of rotation | yes |
Timer | yes |
Timer display | TFT |
Time setting range | 1 - 120 min |
Heating blanket | yes |
Heat output | 250 W |
Heat control | TFT |
Temperature control | PID |
Working temperature | 5 - 80 °C |
Connection for ext. temperature sensor | PT1000 |
Temperature display | yes |
Temperature measuring range | 0 - 100 °C |
Temperature measurement resolution | 0.1 K |
Temperature control accuracy | +/-0.2 °C |
Temperature limit min. | 0 °C |
Temperature limit max. | 110 °C |
pH measurement | yes |
pH control mode | off, auto, profile |
pH value display | TFT |
pH measuring range | 0 - 14 pH |
pH measurement resolution | 0.01 pH |
Accuracy of pH measurement | 0.02 pH |
Conncetion for pO2 sensor | yes |
pO2 control mode | off, manual, auto, profile |
Measurement range pO2 min. | 0 %sat |
Measurement range pO2 max. | 200 %sat |
Resolution pO2 measurement | 0.01 %sat |
Accuracy of pO2 measurement | 0.3 %sat |
Conncetion for turbidity sensor | yes |
Turbidity control mode | Monitoring |
Conncetion for conductivity sensor | yes |
Conductivity control mode | Monitoring |
Conncetion for CO2 sensor | yes |
CO2 control mode | Monitoring |
Gas in connection Air | 6 mm |
Gas out connection Air | 6 mm |
Gas in connection O2 | 6 mm |
Gas out connection O2 | 6 mm |
Gas in connection N2 | 6 mm |
Gas out connection N2 | 6 mm |
Gas in connection CO2 | 6 mm |
Gas out connection CO2 | 6 mm |
Aeration mode | continuous & automatic |
Display flowrate | l/min, cc/min |
Input pressure | 200 - 3000 mbar |
Pressure out min. | 200 mbar |
Pressure out max. | 1000 mbar |
Flowrate Air min. | 0 cc/min |
Flowrate Air max. | 2000 cc/min |
Flowrate O2 min. | 0 cc/min |
Flowrate O2 max. | 2000 cc/min |
Flowrate N2 min. | 0 cc/min |
Flowrate N2 max. | 2000 cc/min |
Flowrate CO2 min. | 0 cc/min |
Flowrate CO2 min. | 2000 cc/min |
Measuring and control accuracy of the gas control units | 2 % fsr |
Peristaltic pump | yes |
Numbers of pump heads | 4 |
Pump head function | Acid, Base, AntiFoam, Level |
Pump head | Watson Marlowe 114DV |
Number of pump tubes | 4 |
Number of pump rollers | 4 |
Flow rate min. | 1.0 ml/min |
Flow rate max. | 270 ml/min |
Pump Speed min. | 100 rpm |
Pump Speed max. | 400 rpm |
Pump reversible direction of rotation | yes |
Pump speed display | TFT |
Inner tube diameter min. | 0.5 mm |
Inner tube diameter max. | 4.8 mm |
Tube outer diamter min. | 4.1 mm |
Tube outer diamter max. | 8.4 mm |
Tube material | Silicone |
Tube hardness | 50 - 60 Shore A |
Dimensions (W x H x D) | 224 x 624 x 405 mm |
Weight | 14 kg |
Permissible ambient temperature | 5 - 40 °C |
Permissible ambient conditions | 80% (up to 31°C), decreasing linearly to max. 50 ( at40°C) |
Protection class according to DIN EN 60529 | IP 20 |
RS 232 interface | yes |
uSB interface | yes |
WiFi Interface | yes |
Voltage | 100 - 240 V |
Frequency | 50/60 Hz |
Power input | 300 W |
Power input max. | 1100 W |
Power input standby | 10 W |
Fuse | 2x T5.0A 250V |